Polar 4 For Windows
Electrochemical simulation and data analysis
DrHuang Pty Ltd
124 Eastern Avenue, Kingsford, Sydney, NSW 2032, Australia
Phone: (61 2) 9662 0516
mailto:
polarography@bigFoot.comCopyright @ 1990-1999
2/12/1999
Contents
1. Introduction
2. Features
3. Menu
4. Input
5. Playing Around
5.1 Running Simulation
5.2 Comparing Curves
5.3 Analyzing Data
5.4 Extracting Parameters by Curve Fitting
5.5 Stripping Voltammetry
6. Frequently Asked Questions (FAQ)
7. References
It analytically and digitally simulates voltammograms (polarograms) on any mechanism at 8 electrode geometry (planar, spherical, semi-spherical, cylindrical, semi-cylindrical, microdisc, thin film, and rotating electrodes) in over 5 techniques (linear sweep and CV, DC, normal pulse, differential pulse, and square wave voltammetries). It outputs current, resistance, conductivity, and surface concentration. It also simulates effects of charge current, resistance, noise, electrolyte, stripping time, stripping potential, etc. User can type in his mechanism.
It analyses any ASCII x-y data for detecting peak location, peak value, semi-derivative, derivative, integral, semi-integral, curve fitting, and separating overlapped peaks.
It shows tip when the user put mouse cursor over a label. The program can separate overlapped voltammograms into individuals, and extract real peak from voltammogram with noise and baseline. It outputs the theoretical peak values, the peak current and potential and current-potential data, which can be imported into other program (e.g. Lotus 123). User can copy-and-paste the voltammogram into his document.
It has been successfully applied to fit experimental polarograms (voltammograms) of In(III), Cd(II), Pb(II), Tl(I), Cr(III), Zn(II), and binuclear copper complex in aqueous and non-aqueous media at mercury, solid metal and non-metal electrodes (specifically the dropping mercury, hanging mercury drop, gold, platinum and glassy carbon electrodes) by various electrochemical techniques (differential pulse, square wave, and pseudo-derivative normal pulse polarographies) [1-5].
It is available from the author or download from my Web site. If you have any question, please read FAQ in its document.
Table 2.1 Features
----------------------------------------------------------------------------- version Public Student Standard Full competitor digital simulation y y y y y analytical simulation y y y y n theoretical peak y y y y n Techniques: LSV, CV y y y y y DC y y y y n normal pulse y y y y n differential pulse y y y y n cyclic DPV y y y y n square wave y y y y n Output: current y y y y y resistance y y y y n conductivity y y y y n surface concentration y y y y y Effect: chemical mechanisms y y y y y noise y y y y y charge current y y y y y resistance y y y y y electrolyte y y y y n preconcentration y y y y n pre-equilibration y y y y y conductivity y y y y n Analysis: derivative y y y y n integral y y y y n semi-derivative y y y y n semi-integral y y y y n manual fit n y y y y auto fit n n y y y manual separate n n n y n auto separate n n n y n Electrode: planar y y y y y (micro)spherical y y y y y (micro)hemispherical y y y y y (micro)cylindrical y y y y y (micro)hemicylindrical y y y y y microdisc y y y y n thin film y y y y y rotating disc y y y y y tip y y y y n import data n y y y y export data n y y y n ----------------------------------------------------------------------------- note: y = yes, n = no. price may be changed.
File menu
Input menu
Run menu
Display menu
Analysis menu
Help menu
Some menus will be activated only after you click the Simulate submenu or load data because they require data.
Chapter 4
Input
4.1 Techniques window
1) Linear sweep and cyclic voltammetry
2) DC voltammetry
3) Normal pulse voltammetry
4) Differential pulse voltammetry
5) Square wave voltammetry
The shapes of DC and normal pulse polarogram are S-shape. The shapes of differential pulse and square wave voltammograms usually are peak-shape. But there is effect of the DC term on differential pulse voltammogram.
4.2 Mechanism window
User can type in his mechanism in Digital Simulation section. In order to faster computation, you should type in reactants only without products if chemical reaction is irreversible.
Instrumental Parameters Section:
E start: starting potential (V).
E end: ending potential (V).
E step: step potential (V).
v: scan rate (V/s). For square wave voltammetry, v=E step/t pulse.
E pulse: pulse potential (V).
T: temperature (
Scan:
Electrode Section:
Planar: planar electrode.
(Micro)Spherical: spherical electrode or micro spherical electrode.
(Micro)Hemispherical: hemispherical electrode or micro hemispherical electrode.
(Micro)Cylindric: cylindrical electrode or micro cylindrical electrode.
Microdisc: microdisc electrode, radius <1e-4 cm
Thin film: thin film electrode
Rotating disc: rotating disc electrode
Area: electrode area (cm2).
Radius: electrode radius (cm).
Length: electrode length for cylindrical electrode or micro cylindrical electrode, or mercury film thickness for stripping voltammetry (cm).
Preconcentration Section:
E pre: preconcentration potential (V).
R stir: stirring rate (rpm). Stirring solution
t pre: preconcentration time (s).
t pre const: preconcentration time constant (/s).
P const: electrode constant. It only related to electrode.
4.3 Chemicals Window
Species Section:
D: diffusion coefficient (cm2/s).
C anal: analytical concentration (M).
C init: initial concentration for simulation (M).
C fitted: fitted value of concentration (M).
C min: minimum concentration for fitting (M).
C max: maximum concentration for fitting (M).
Heterogeneous Reaction Section:
ks: heterogeneous standard rate constant (cm/s).
Homogeneous Reaction Section:
kf: forward chemical reaction rate constant.
kb: backward chemical reaction rate constant.
Kq: chemical equilibrium constant, Kq = kf/kb.
A simplest way to play simulation is just to click the Simulate submenu under the Run menu. It uses the default values to simulate a linear sweep voltammogram.
Notice that some menu (e.g. the Display menu and the Analysis menu) will be activated only after run simulation or load data because they require data.
5.1.1 Effect of Electrode Size - Microelectrode
Simulation technology for microelectrode is the same as for macro electrode, but the electrode size is very small, e.g. electrode radius is 1e-4 cm. A shape of voltammogram will be changed. Note that the planar electrode geometry is not available for microelectrode.
After run first simulation, click the Display menu, and click the Option submenu. Select the Overlap choice, then run second simulation.
Semi-derivative is useful for CV. It can change a shape of CV into symmetric peak if CV is reversible.
5.4 Extracting Parameters by Curve Fitting
5.4.1 Fitting to Simulation Curve
In order to extract kinetic parameters, you can fit a simulation curve to another simulated or experimental curve. You should manual fit before auto fit. The manual fit shows how well your initial guess values work. It can retrieve any of 20 parameters (concentration C, standard electrode potential E°, and the heterogeneous standard rate constant ks) from voltammogram by curve fitting. If it diverged, you should change their initial values, then try again.
e.g. run simulation with all default values, then change the C value from 1e-3 to 2e-3 in the Species section, click the Auto Fit menu. You will see the fitted value of 0.001 in the C fitted field next to the C text field.
5.4.2 Fitting to Experimental Curve
It is similar to fit simulated curve. But you should input your experimental values of E start, E end, E step, etc. into the Experimental section. Polar requires data are in SI unit and first peak is positive value. If your experimental data are not, please convert your experimental data.
Select the Preconcentration in the Experiential Parameters window. Change the preconcentration potential value in the E pre text field, and preconcentration time in the t pre text field. The preconcentration potential value usually is -0.2/n V to specie’s standard electrode potential. The preconcentration time usually is a number of minutes. You should enter your electrode constant into the P cont text field, and your mercury film thickness into the Length field in the Electrode section of the Experimental window if you use a planar mercury film electrode.
Chapter 6
Frequently Asked Questions (FAQ)
Q: Which platforms can Polar run on?
The 32-bit version needs Microsoft Visual Basic 6 runtime DLL files (e.g. msvbvm60.dll, comdlg32.ocx) in the same directory as Polar or in the directory \windows\system for Windows 3.11 or 95, or in the directory \winnt\system32 for Windows NT.
The 16-bit version needs Microsoft Visual Basic 4 runtime DLL files (e.g. vb40016.dll and oc25.dll) in the same directory as Polar or in the directory \windows\system for Windows 3.1, or in the directory \winnt\system for Windows NT.
Q: Where can I download these dll?
A: Microsoft Visual Basic 6 runtime DLL files are
from http://www.simtel.net/simtel.net/win95/dll.html, where msvbvm60.dll is inside simvb6-3.zip. Microsoft Visual Basic 4 16-bit runtime DLL files are from http://www.simtel.net/simtel.net/win3/dll.html.Q: When I click the Simulate menu, I got error: No data.
Q: I cannot save a file.
A: You miss the Microsoft Visual Basic 6 runtime DLL file comdlg32.ocx.
Q: When I installed to run setup.exe, an error occured:
Q: Still have install problem?
Cd \windows\system
Regsvr32 msvbvm60.dll
then start Polar.
Q: Why are some menus grey?
Q: I cannot see any chemical reaction in Public version. Is this part of the program not finished yet or is it only available in the registered version?
Q: Does it include my mechanism?
Q: Can it fit data by curve fitting?
Q: Can I change graph into other program Lotus 123 or Excel?
Q: Some submenus semi-derivative, semi-integral, derivative, and integral, seem to not work sometime. How can I do?
Q: How much does registration cost?
Q: How can I get registered version?
Q: What are difference among Public, Standard and Professional, and Full versions?
Q: When I run the SWV with default conditions as a digital simulation, it does not appear to give the correct curve. Why?
Q: Is it possible to click on a point and then have displayed both the current and potential for the point? Also, does it have any zoom features?
Q: How does it compare to competitors?
[1] W. Huang, T. Henderson, A.M. Bond and K.B. Oldham, Curve fitting to resolve overlapping voltammetric peaks: model and examples, Anal. Chim. Acta, 1995, 304, 1-15.
[2] W. Huang and B. Hibbert, Computers & Chem., 1995, 19(4), 433.
[3] W. Huang and B. Hibbert, Computers & Chem., 1995, 19(4), 435.
[4] W. Huang and B. Hibbert, Polar 2.0 for Windows: simulator of voltammogram, Chem. in Aus., 1996, 131.
[5] J. Mo, P. Cai, W. Huang and F. Yun, Theory and application on multiple semidifferential electrochemical stripping analysis with thin mercury film formed in situ, Acta Chimica Sinica, 1984, 42(6), 556-561, CA 101: 162712.